Jual Midi LOGGER GL240 GRAPHTEC
| Terakhir Diupdate | 2026-04-18 |
| Negara Asal | Amerika |
| Pembelian Minimum | 1 |
| Harga | |
Butuh Penawaran Harga, bantuan atau Konsultasi Produk?
Jual Midi LOGGER GL240 GRAPHTEC
Input/Output cable for GL (option B-513) is required to connect the signal.
Input signal;
- Voltage range: Up to 24V (common ground)
- Signal type: Voltage, Open collector, Contact (relay)
- Threshold: Approx. + 2.5 V (Hysteresis: Approx. 0.5V (2.5V to 3V))
Output signal: Open collector (pull-up to 5V by 10kΩ resistor)
- Voltage: Max. 30V
- Current: Max. 0.5A
- Collector dissipation: Max. 0.2W
Minimum interval varies by number of channels used.Output port can be specified in each input channel.
The built-in Flash memory is available for units with serial numbers C6043XXX or later. Please contact your local representative for more information.SD memory card cannot be used on SD card slot while the wireless LAN unit (opton B-568) is used.
Size of the capture data will be limited to 1/3 of available memory.Display mode is switched every time the dedicated key is pressed. In magnified digital value mode, the displayed channel number can be specified. In the waveform display mode, the changing of the time scale will be effective from the point of the next displayed data.
Rating under maximum power consumption using the AC adapter, with LCD display on, and battery pack being
charged.Excludes AC adapter and battery pack.
The terminal "b" for using the RTD is connected each other across all channels.Subject to the following conditions:
- Room temperature is 23 oC ± 5 oC.
- When 30 minutes or more have elapsed after power was turned on.
- Filter is set to 10.
- Sampling rate is set to 1 sec, using 10-channel.
- GND terminal is connected to ground.
Wire size of thermocouple used is 0.32mm diameter in the T or K type and 0.65mm diameter in other types.
+ Call